Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 891
Chapter 9
ERR-SUB-A-003.fm
122/071
DEBUG
ONLY
COMPRESSION
INVALID
COMPRESSION_PE
DAL_PRESS
Occurs when the compression pedal is depressed.
Ignore this error. If the problem persists, contact your
OLC.
122/072
DEBUG
ONLY
COMPRESSION
INVALID
DECOMPRESSION
_PEDAL_PRESS
Occurs when the decompression pedal is
depressed.
Ignore this error. If the problem persists, contact your
OLC.
122/073
DEBUG
ONLY
COMPRESSION
INVALID
COLLIMATOR_BUT
TON_PRESSED
Occurs when the collimator button is depressed.
Ignore this error. If the problem persists, contact your
OLC.
123/001 ADVISORY
LCD: A24 Maximum
compression force is
reached
MAX COMPR
FORCE ALREADY
Occurs when an operator presses the compression
footswitch whereas the maximum compression force
is already reached.
Operator error. No service action required.
123/002 ADVISORY
LCD: A05 Compres-
sion upper position is
reached
COMPR UP LIMIT
ALREADY
Occurs when an operator presses the decompres-
sion footswitch whereas the upper software limit for
the compression paddle positioning is already
reached.
Operator error. No service action required.
123/003 ADVISORY
LCD: A23 Compres-
sion lower position is
reached
COMPR LOW LIMIT
ALREADY
Occurs when an operator presses the compression
footswitch whereas the lower software limit for the
compression paddle positioning is already reached.
This can happen with no Breast Support and no
Bucky.
Operator error. No service action required.
123/004 ADVISORY COMPR NEGATIVE
LIMIT REACHED
Negative force limit reached.
Occurs when the decompression motion is ongoing/
requested and a negative force limit is reached
(-3 daN).
Operator error. No service action required.
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action