ERR SUB A003 - Gantry Error Messages
3-6
Error Codes from 150/yyy: PDU
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card
152/000
DEBUG
ONLY
PDU 230V AC mains
is out of the spec lim-
its
The 230 V AC mains is out of the specified limits.
See
JC DIAG A1500 AC MAINS FAILURE
to fix the problem.
A1500
152/001
WARNING
12VDC Generator
voltage off
The 12 V DC signal from Generator Supply Com-
mand board to PDU board is not present.
See
JC DIAG A1501 12V DC Generator Voltage Off
to fix the problem.
A1501
152/002
DEBUG
ONLY
48V DC channel
error
The AC/DC module is turned on but no 48 V DC is
detected.
See
JC DIAG A1502 ACDC MODULE POWER
to fix the problem.
A1502
152/003
WARNING
Power Fail
The power on the AC/DC module has failed.
See
JC DIAG A1508 AC/DC Module Reports Power
to fix the problem.
A1508
152/004
DEBUG
ONLY
Fan Fail
The fan on the AC/DC module has failed.
See
JC DIAG A1509 AC/DC Module Reports Fan
to fix the problem.
A1509
152/005
WARNING
Over Temperature
The temperature of the AC/DC module has
exceeded the specified limits.
See
JC DIAG A1510 AC/DC Module Reports Over
A1510
152/006
DEBUG
ONLY
Internal 12VDC volt-
age is out of spec
limits
The internal (regulated) 12 V DC voltage is out of the
specified limits.
Ignore this error. If the error persists, contact your
OLC.
152/007
DEBUG
ONLY
Total current con-
sumption is too high
The total current consumption of PDU is too high (i.e.
greater than 40 A).
See
JC DIAG A1503 OVERCURRENT FAILURE
to fix the problem.
A1503
152/008
DEBUG
ONLY
GND offset. Level of
PDU ground is too
high
The level of PDU ground is too high (possible noise
is present).
Check the Shunt R101 and cables W109, W111, and
W113.
If the problem persists, change PDU board (see