JC-TSG-A-016.fm
Ensure that the clutch cables are correctly positioned, and secured to the spacer of Rotation Board
PL202 by a cable tie as indicated in illustration 3.
ILLUSTRATION 3 - RECOMMENDED ROTATION CLUTCH CABLE ROUTING
Pass a cable tie (A) through the
hole in the curtain metal support
to secure the cables behind the
curtain.
The free length between the two
cable ties (A and B) is approx. 60 cm
(24 inch).
Use some adhesive tape to secure
the two cables together in this area.
Do not use cable ties; they can
catch on other items during
movement.
Use a cable tie (B) to
secure the cables at
this point.
When installing the covers, place these cables inside the
lower cover with the keypad cable. The cables must
move freely; check that they cannot catch on anything.
Pass a cable tie (A) through the
hole in the curtain metal support
to secure the cables behind the
curtain.
Correct cable routing around top