Chapter 6
Page no. 514
JC-PHY-A-042.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A042 - Remove/Reinstall Generator Covers
6
REMOVAL AND REINSTALLATION OF COVERS
6-1
Remove Generator Covers
1. Remove all power from the Senographe.
2. Remove the four plastic blanking disks (1) (slim flat blade) in each corner of the Generator top cover
to expose the screws that secure the Generator top cover.
3. Remove the four allen screws (2) that secure the Generator top cover (5.5 mm allen wrench).
1
1
1
1
2
2
2
2