GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 919
Chapter 9
ERR-SUB-A-003.fm
162/009
HALT
STEREOTIX HARD-
WARE DETECTION
IN FAILURE
The Gantry CPU failed to detect the stereotix hard-
ware.
Reboot the system from the X-Ray Console. If the
problem persists contact your OLC.
162/010
INHIBIT
STEREOTIX CAN
NOT ENTER
READY STATE
The stereotix node can not switch to the READY
state.
Reboot the system from the X-Ray Console. If the
problem persists contact your OLC.
162/011
INHIBIT
STEREOTIX CAN
NOT ENTER PUNC-
TURE STATE
The stereotix node can not switch to the PUNCTURE
state.
Reboot the system from the X-Ray Console. If the
problem persists contact your OLC.
162/012
WARNING
STEREOTIX TAR-
GET LIST IS FULL
The stereotix node's target list is FULL.
Reboot the system from the X-Ray Console. If the
problem persists contact your OLC.
162/013
HALT
STEREOTIX NODE
NOT RESPONDING
Heartbeat signal was not sent by the stereotix node.
The stereotix node sends a Heartbeat signal every
second to Gantry CPU. If not, the Gantry CPU cre-
ates this error. This error occurs when an operator
did not disconnect the stereotix within 10s after
releasing the latches (time out).
Operator error. After releasing the latches, operator
has 10s to disconnect the stereotix. Reboot the sys-
tem from the console then redo the operation into the
allocated timestamp.
162/014
HALT
INTERNAL SOFT-
WARE ERROR
The Gantry CPU received an unknown error code
from the stereotix node.
Reboot the system from the X-Ray Console. If the
problem persists contact your OLC.
162/015 ADVISORY LCD: A19 Stereotac-
tic Positioner is not
ready
The stereotix positioner is not ready.
Operator error. No service recommended actions.
162/016 ADVISORY LCD: A12 Move tube
head to center posi-
tion
Move to angulation 0°
Operator error. No service recommended actions.
162/017 ADVISORY
LCD: A16 Rotation
limit for Stereotaxy is
reached
Stereotactic rotation limit reached.
Operator error. No service recommended actions.
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card