Job Card ELE A020 - Declare and Set Printers (Filming Devices)
•
Performing physicist: maximum number of characters allowed with "W" letter in upper case (64
letters).
•
Operator: maximum number of characters allowed with "W" letter in upper case (64 letters).
2. Put the ACR phantom on the Bucky and perform one exposure in manual mode with the following
settings: Mo/Mo 22kV 4mAs, with a field of view 19x23.
3. Perform a manual print with the printing annotations used by the site of this image.
Note:
If many print annotations are used at the site, each of them has to be tested. If the printer contained
several film sizes, each film size has to be tested.
Note:
To change the printing annotations, use the Annotations window in Medical Preferences. Refer to
the Browser chapter in the
Senographe DS Operator Manual
for access to this window.
No overlap of the annotation must be seen on the printed film, and there must be all 64 Ws visible in
the patient name, patient ID, performing physicist and operator with no interference from other anno-
tations (see Illustration 16). If an overlap of the site name/site address is seen on the printed film, you
need to reduce the number of letters put for the site name/site address. Refer to
for editing the site name/site address.