ERR SUB A003 - Gantry Error Messages
Page no. 885
Chapter 9
ERR-SUB-A-003.fm
122/026
HALT
COMPRESSION
NODE PARITY
ERROR
Occurs when the parity bit does not set correctly on
the on-board bus between the CPLD and the MCU.
1. Change the Compression Board.
2. Perform a calibration of the Compression Force
Sensor.
3. Perform a Compression Thickness calibration.
D/R A250
CAL A048
CAL A050
122/027
N/A
STOP LINE WAS
ACTIVATED BY
CPU
The main CPU detected an error and activated the
stop motion line or did not deactivated it at all.
Ignore this error. If the problem persists, contact your
OLC.
122/028
HALT
LOW VOLTAGE
ERROR ON COM-
PRESSION NODE
If the voltage of the power supply falls below 24 V
the node sends this error.
1. Change the Compression Board.
2. Perform a calibration of the Compression Force
Sensor.
3. Perform a Compression Thickness calibration.
D/R A250
CAL A048
CAL A050
122/029
HALT
COMPRESSION
READ TIMEOUT
ERROR
If the MCU does not react on the IRQ coming from
the CPLD before 15 ms, the node sends this error.
1. Change the Compression Board.
2. Perform a calibration of the Compression Force
Sensor.
3. Perform a Compression Thickness calibration.
D/R A250
CAL A048
CAL A050
122/030
HALT
COMPRESSION
CPLD MEMORY
ERROR
Test data from the CPLD is periodically read to
ensure the proper communication. If this error does
not match the node sends this error.
1. Change the Compression Board.
2. Perform a calibration of the Compression Force
Sensor.
3. Perform a Compression Thickness calibration.
D/R A250
CAL A048
CAL A050