GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card DIAG A001 - AWS Sun-VTS Diagnostics
Page no. 979
Chapter 9
JC-DIAG-A-001.fm
5
PHYSICAL MEMORY TESTS
•
Type of test:
SUN VTS
On V1/V2 Control Stations, from a command window, enter:
/opt/SUNWvts/bin/pmem –v -f -o size=100
On V3 Control Stations, from a command window, enter:
/opt/SUNWvts/bin/pmemtest –v -f -o size=100
6
VIRTUAL MEMORY TESTS
6-1
Short Test
•
Type of test:
SUN VTS
On V1/V2 Control Stations, from a command window, enter:
/opt/SUNWvts/bin/vmem –v -f -o amount=100
On V3 Control Stations, from a command window, enter:
/opt/SUNWvts/bin/vmemtest –v -f -o amount=100
6-2
Long Test
•
Type of test:
SUN VTS
On V1/V2 Control Stations, from a command window, enter:
/opt/SUNWvts/bin/vmem –v –f -o reserve=100
On V3 Control Stations, from a command window, enter:
/opt/SUNWvts/bin/vmemtest –v –f -o reserve=100
7
VIDEO BOARD TESTS
•
Type of test:
SUN VTS
Note:
Video board tests are only applicable to V1/V2 Control Stations.
On V1/V2 Control Stations, from a command window, enter:
/opt/SUNWvts/bin/m64test
8
FLOPPY DRIVE TESTS
•
Type of test:
SUN VTS
Note:
Floppy drive tests are only applicable to V1/V2 Control Stations.
1. From a command window, log in as root as follows:
a. At the ADS prompt enter:
su