Chapter 11
Page no. 1794
JC-PM-A-300.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card PM A300 - 21-inch 3MP Monitor Luminance Checks
4
APPLICABILITY
The Job Card must be applied during the planned maintenance visit on systems containing the optional
21-inch 3MP Monitor (P/N 5350390).
5
SAFETY PRECAUTIONS
None
6
PREREQUISITES
The 21-inch 3MP Monitor has already been installed and configured to use
1 - DICOM
mode and the
maximum luminance is 400 cd/m² according to
ELE A008 - Configuring the 21-inch 3MP Monitor
7
PROCEDURE
7-1
Install Test Images
1. Insert the Test Pattern CD-ROM into the DVD-ROM drive of the ADS Computer.
2. In the Browser, click the
CDR read/write
icon and select
Query
from the menu that appears.
A
Media Browser:CDR
window appears.
3. From the
Media Browser:CDR
window:
a. Select the
AAPM Test Patterns
patient.
b. Select the
1K-LUMIN
Study ID.
c. Select both the
2
and
4
Series (use the
Control
key to select both Series).
d. Click the
Disk
icon under the Series list. From the
Restore
dialog box that appears, click
OK
.
e. The test patterns from the Test Pattern CD-ROM are transferred to the patient list on the ADS
Computer. During this transfer time, an animated data transfer icon towards of the top left of the
Media Browser:CDR
window appears. When the transfer of the test patterns is complete, the data
transfer icon disappears.
d
a
b
c
e