ERR SUB A003 - Gantry Error Messages
164/004
INHIBIT
COMMUNICATION
ERROR WITH COL-
LIMATOR NODE
The Gantry CPU was unable to load the collimator
parameters.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
164/005
INHIBIT
INTERNAL SOFT-
WARE ERROR
The Gantry CPU received an unknown error code
from the collimator node.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
164/006
INHIBIT
COLLIMATOR CON-
TROL ERROR
The collimator node did not start. At power-up, the
collimator node sends an acknowledgment (which
can be different from the first heartbeat signal) to
Gantry CPU. If the collimator node does not send an
acknowledgment, the Gantry CPU creates this error.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
164/008
INHIBIT
COLLIMATOR
NODE NOT
RESPONDING
The collimator node did not send the Heartbeat sig-
nal.
Between two motions, the collimator node sends a
Heartbeat signal every second to Gantry CPU. If not,
the Gantry CPU creates this error.
During lift motion, the lift collimator sends an actual
value message regularly in addition to the heartbeat.
If not, the Gantry CPU creates this error.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
165/000
INHIBIT
Internal software
error
There is an invalid serial port configuration, and
therefore no communication is possible between the
Gantry CPU and the Generator.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
165/001
INHIBIT
Internal software
error
There is a hardware error with the COM port, and
therefore no communication is possible between the
Gantry CPU and the Generator.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
165/002
INHIBIT
Internal software
error
Received message number does not match.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card