GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 875
Chapter 9
ERR-SUB-A-003.fm
112/045
HALT
ROTATION OPTI-
CAL FORK DEFECT
At power-up, the rotation node shuts down the LED
of the left optical fork, and checks that the photore-
sistor shuts down. The optical forks are hardware
limits hence they are never obstructed.
Refer to
JC DIAG A1118 ROTATION CONTROL
to fix the problem.
A1118
112/046
DEBUG
ONLY
ROTATION INVALID
USER INTERFACE
ERROR
Ignore this error. If the problem persists, contact your
OLC.
112/047
DEBUG
ONLY
ROTATION INVALID
LIFT UP SLOW
BUTTON ERROR
Ignore this error. If the problem persists, contact your
OLC.
112/048
DEBUG
ONLY
ROTATION INVALID
UP FAST BUTTON
ERROR
Ignore this error. If the problem persists, contact your
OLC.
112/049
DEBUG
ONLY
ROTATION INVALID
LIFT DOWN SLOW
ERROR
Ignore this error. If the problem persists, contact your
OLC.
112/050
DEBUG
ONLY
ROTATION INVALID
LIFT DOWN FAST
BUTTON ERROR
Ignore this error. If the problem persists, contact your
OLC.
112/051
DEBUG
ONLY
ROTATION INVALID
ROT LEFT SLOW
BUTTON ERROR
Ignore this error. If the problem persists, contact your
OLC.
112/052
DEBUG
ONLY
ROTATION INVALID
ROT LEFT FAST
BUTTON ERROR
Ignore this error. If the problem persists, contact your
OLC.
112/053
DEBUG
ONLY
ROTATION INVALID
ROT RIGHT SLOW
BUTTON ERROR
Ignore this error. If the problem persists, contact your
OLC.
112/054
DEBUG
ONLY
ROTATION INVALID
ROT RIGHT FAST
BUTTON ERROR
Ignore this error. If the problem persists, contact your
OLC.
112/055
DEBUG
ONLY
ROTATION INVALID
MLO LEFT BUT-
TON ERROR
Ignore this error. If the problem persists, contact your
OLC.
112/056
DEBUG
ONLY
ROTATION INVALID
MLO RIGHT BUT-
TON ERROR
Ignore this error. If the problem persists, contact your
OLC.
112/057
DEBUG
ONLY
ROTATION INVALID
CC BUTTON
ERROR
Ignore this error. If the problem persists, contact your
OLC.
112/058
DEBUG
ONLY
ROTATION INVALID
COLLIMATOR BUT-
TON ERROR
Ignore this error. If the problem persists, contact your
OLC.
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card