Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A210 - Gantry cover replacement guide
Page no. 1209
Chapter 9
JC-DR-A-210.fm
3-7-1
Detector Covers - non-available covers
•
If the reference of the defective part appears in the following table, it cannot be replaced individually.
A complete set of covers must be ordered and installed:
-
Order a set of covers as listed below:
•
The Detector must be modified to accept the new covers. Refer to
•
The new Detector/Positioner Interface Plate reference 2408213, together with the new Quick Fix
Assembly, reference 2411952, must be installed in place of the existing Interface Plate. Refer to
•
After modifying the detector and changing the interface plate, install the new covers using the infor-
mation given in
Job Card PHY A044 - Remove/Reinstall Gantry Covers
,
3-7-2
Detector Covers - available covers
If the reference of the defective part appears in the following table, it can be replaced individually.
The table indicates the necessary part(s) to be ordered
•
Install the new cover(s) using the information given in
Job Card PHY A044 - Remove/Reinstall Gan-
Reference
Description
2375024
2375024-3
Detector Cover Upper
2375025
2375025-3
Detector Cover Lower
Reference
Description
Quantity
2375024-2
Detector Cover Upper
1
2375025-2
Detector Cover Lower
1
2411952
Quick Fix Assy for Detector Covers
1
2408213
Interface Plate Equipped
1
If the defective part is:
Order the following parts:
Reference
Description
Quantity
2375024-2
Detector Cover Upper
2375024-2
Detector Cover Upper
1
2375025-2
Detector Cover Lower
2375025-2
Detector Cover Lower
1