Pre-Installation System Requirements
Page no. 355
Chapter 4
Chap-Pre-Requirements.fm
10-3-2-2. Anchoring Inserts Not Provided with the Senographe System (for seismic areas)
The following tables provide the recommended anchoring inserts seismic areas. Due to the differing
availability of Hilti bolts in different contentinents, the following tables provide different bolt recommenda-
tions. The minimum floor thickness required varies according to the bolts used.
USA - Hilti Kwick Bolt III
EUROPE - Hilti HSA Bolts
Anchoring holes in the floor
Gantry baseplate
(see Illustration 15)
Gantry baseplate addi-
tional anchoring holes for
seismic areas
Control station base-
plate
Number of holes in the plate
3
2
4
Diameter of the hole in the
plate
11 mm
11 mm
14 mm
Recommended inserts
Hilti Kwik Bolt III 3/8"
Hilti Kwik Bolt III 3/8"
Hilti Kwik Bolt III 3/8"
Hole diameter in the floor
9.5 mm (3/8”)
9.5 mm (3/8”)
9.5 mm (3/8”)
Hole depth in the floor
64 mm (2 ½”)
64 mm (2 ½”)
64 mm (2 ½”)
Minimum floor thickness
95 mm (3 ¾”)
95 mm (3 ¾”)
95 mm (3 ¾”)
Maximum bolt load pull ten-
sion (at each bolt)
490 daN
490 daN
490 daN
Recommended tightening
torque
34 Nm
34 Nm
34 Nm
Anchoring holes in the floor
Gantry baseplate
(see Illustration 15)
Gantry baseplate addi-
tional anchoring holes for
seismic areas
Control station base-
plate
(see Illustration 16)
Number of holes in the plate
3
2
4
Diameter of the hole in the
plate
11 mm
11 mm
14 mm
Recommended inserts
HSA M10 x 90 - 20 / 25
HSA M10 x 90 - 20 / 25
HSA M10 x 90 - 20 / 25
Hole diameter in the floor
10 mm
10 mm
10 mm
Hole depth in the floor
70 mm
70 mm
70 mm
Maximum bolt load pull ten-
sion (at each bolt)
490 daN
490 daN
490 daN
Minimum floor thickness
105 mm
105 mm
105 mm
Recommended tightening
torque
34 Nm
34 Nm
34 Nm