Chapter 9
Page no. 1230
JC-DR-A-222.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A222 - Clutch and Cable
The cables from the clutch to the Rotation Board PL202 must be secured as shown:
6-4
Check Rotation Movement
Move the compression paddle out of compression. Then check that all four sets of keypad control but-
tons on the arm and the tube head are operational (check each button for arm rotation movement).
7
COMPLETION
None
Pass a cable tie (A) through the
hole in the curtain metal support
to secure the cables behind the
curtain.
The free length between the two
cable ties (A and B) is approx. 60 cm
(24 inch).
Use some adhesive tape to secure
the two cables together in this area.
Do not use cable ties; they can
catch on other items during
movement.
Use a cable tie (B) to
secure the cables at
this point.
When installing the covers, place these cables inside the
lower cover with the keypad cable. The cables must