Chapter 9
Page no. 910
ERR-SUB-A-003.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
3-7
Error Codes from 160/yyy: CPU
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card
160/000
HALT
INTERNAL SOFT-
WARE ERROR
One of the following out of memory errors occurred:
- functional device creation fails
- configuration interface creation fails
- HAL device creation fails
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
160/001
HALT
INTERNAL SOFT-
WARE ERROR
A Common Object Request Broker Architecture
(CORBA) extraction failed.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
160/002
HALT
INTERNAL SOFT-
WARE ERROR
An unknown attribute change notification, HAL sig-
nal, or Atlas event was received.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
160/003
HALT
INTERNAL SOFT-
WARE ERROR
The Gantry CPU failed to read the configuration file.
1. Reboot the system from the X-ray Console.
2. If the problem persists change the CPU Board
and re-calibrate the Collimator afterwards.
3. If the problem persists, contact your OLC.
D/R A190
CAL A029
ELE A028
160/004
HALT
INTERNAL SOFT-
WARE ERROR
The Gantry CPU failed to get the configuration data.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
160/005
HALT
INTERNAL SOFT-
WARE ERROR
An Atlas event, attribute change, or HAL signal sub-
scription has failed.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
160/006
HALT
INTERNAL SOFT-
WARE ERROR
The Gantry CPU was unable to access the state
machine file.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
160/007
HALT
INTERNAL SOFT-
WARE ERROR
The Gantry CPU failed to get an attribute.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.