ERR SUB A003 - Gantry Error Messages
Page no. 869
Chapter 9
ERR-SUB-A-003.fm
112/014
HALT
ROTATION OVER-
CURRENT
Occurs when an integrated value of the current is too
high, that is when the current is too high during too
much time.
Refer to
JC DIAG A1110 ROTATION CONTROL
to fix the problem.
A1110
112/015
DEBUG
ONLY
ROTATION SOFT-
WARE ERROR
Occurs when the node is currently performing a
movement and the Gantry CPU sends a command
for another movement while the current movement is
not finished. This is a CPU problem because CPU
did not detect the current movement.
Should not occur in the field. It could be a disconnect
between the CPU soft release and the Node soft
release. If the problem persists, contact your OLC.
112/016
HALT
ROTATION NODE
FW INCOMPATIBIL-
ITY
Occurs when the hardware configuration (CPLD) of
the rotation node, the DSP, or the MCU firmware not
match.
Change the Rotation board then perform a Rotation
Calibration.
D/R A221
CAL A021
112/017
HALT
ROTATION NODE
PARITY ERROR
Occurs when the parity bit does not set correctly on
the on-board bus between the CPLD and the MCU
Change the Rotation board then perform Rotation
Calibration.
D/R A221
CAL A021
112/018
N/A
STOP LINE WAS
ACTIVATED
BECAUSE OF
UNCONTROLLED
MOTION
An uncontrolled motion is detected by the DSP. The
rotation node sends this error and activates the
MCU_STOP_MOTION line which leads to 0x500A
error.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
112/019
N/A
STOP LINE WAS
ACTIVATED BY
CPU
The stop motion line is active. The main CPU
detected an error and activated the stop motion line
or did not deactivate it at all.
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card