Chapter 7
Page no. 606
JC-ELE-A-020.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A020 - Declare and Set Printers (Filming Devices)
7-3-4
DryView 6850 DICOM Printer
•
DICOM printer label: Any name you wish to see appearing on the Film Composer, e.g.,
Carestream 6850.
•
Hostname: Use printer Hostname if any, or a name unique to your Network, e.g., DV6850DRE.
•
Network address: A unique IP address given by the hospital Network Administrator), e.g.,
3.249.15.142.
•
Application Entity Title (AET):
Must be
DV6850
•
Port number:
Must be
5040
•
Layouts:
Must be
1 x 1
, 1 x 2
, 2 x 1
, or 2 x 2
•
Film size:
Must be
8 in x 10 in W = 4924 H = 6224
10 in x 12 in W = 6224 H = 7526
•
Pixel size
39.08
•
Configuration information
Must be
LUT=gsdfaa0p.w8u,3
•
Density
Must be
Min: 20 Max: 360 for DVM film type
Max: 415 for DVM+ film type
•
Magnification type:
Must be
Cubic
•
Smoothing factor:
Must be
15
•
Trim:
Must be
NO
•
Printer Memory Size
40 (default value)
Note:
To match the new AFSSAPS requirements (in France) for Ambient Light=1 & Illumination=4000,
do not set Review Room into Review Room Management, and check with the printer manufacturer
that the printer is correctly configured.