Chapter 9
Page no. 1574
JC-DR-A-406.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A406 - Generator Command Board 400-PL1
TABLE 1 - JP1 AND JP2 JUMPER SETTINGS
0 = No jumper
1 = Jumper present
6. Ensure that all the Test Point (TP) connectors (labeled TP1, ... etc.) on the Generator Command
board are open.
7. Switch on the electrical power supply, both from the Mains Distribution Panel in the room and on the
rotary switch in the Generator cabinet (position I).
8. Switch on the Senographe system electrical power (i.e. Generator, Gantry, and Control Station).
9. Wait until the Gantry boot is complete, and check that no error has been reported.
If the Generator Command board is operating correctly, the LED states on the Generator Command
board are as follows:
10. Reinstall the Generator covers (see
Job Card PHY A042 - Remove/Reinstall Generator Covers
JP1
(PREP)
JP2
(RADIOGRAPHY)
Door closed
Door opened
1
0
0
PREP and RADIOGRAPHY
enabled
PREP and RADIOGRAPHY
disabled
2
0
1
PREP and RADIOGRAPHY
enabled
PREP and RADIOGRAPHY
disabled.
NOTE:
if the door opens during
an exposure, the exposure is not
aborted and completes normally
)
3
1
0
PREP and RADIOGRAPHY
enabled
PREP enabled
RADIOGRAPHY disabled
4
1
1
PREP and RADIOGRAPHY
enabled
PREP and RADIOGRAPHY
enabled
•
green DS1 LED is OFF
•
red DS8 LED is OFF
•
yellow DS2 LED is OFF
•
green DS9 LED ON
•
green DS3 LED is OFF
•
yellow DS10 LED is OFF
•
red DS4 LED is ON
•
green DS11 LED ON
•
yellow DS5 LED is OFF
•
red DS12 LED is OFF
•
green DS6 LED is ON
•
red DS13 LED is OFF
•
yellow DS7 LED ON