Chapter 7
Page no. 790
JC-ELE-A-065.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A065 - Trackball Configuration
6. To check that buttons 1 and 6 are correctly set for
right click
:
With the cursor in the
Console
window, press the keyboard
Control
key.
Keep the
Control
key pressed while you do each of the following checks:
•
Press button 1
:
if the
VT Fonts Menu
does not appear, repeat the setup procedure for button 1.
•
Release button 1
:
if the
VT Fonts Menu
does not close, repeat the setup procedure for button 1.
•
Press button 6
:
if the
VT Fonts Menu
does not appear, repeat the setup procedure for button 6.
•
Release button 6
:
if the
VT Fonts Menu
does not close, repeat the setup procedure for button 6.