Chapter 7
Page no. 550
JC-ELE-A-008.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A008 - Configuring the 21-inch 3MP Monitor
4
SAFETY PRECAUTIONS
None
5
APPLICABILITY
This Job Card is only applicable to Senographe systems running ADS 53.40 and above, which have
optional the 21-inch 3 MP Monitor installed. The 21-inch 3 MP Monitor is only available for V3 and V4
Control Stations.
6
PREREQUISITES
The 21-inch 3 MP Monitor has been physically connected during the physical installation phase of the
Senographe system.
Note:
Ensure that the monitor DVI cable is connected to the left hand side DVI connector on the back of
the ADS computer.
7
PROCEDURE
7-1
Resolution Configuration
The steps you undertake depend on whether you have a V3 or V4 Control Station, as summarized in the
following steering guide.
7-1-1
Resolution Configuration on V3 Control Stations
1. Reboot the ADS Workstation to ensure that you are at the first
Login:
prompt.
A series of system initialization messages appears on the screen, then the
Login:
prompt appears.
2. Login as the root user as follows:
a. At the
Login:
prompt, enter:
root
b. At the
Password:
prompt, enter:
operator
A command prompt appears for the root user.
3. At command prompt, launch the Desktop Manager Configuration Tool, by entering:
kdmconfig
Perform
kdmconfig
to set monitor size
and resolution - see
olution Configuration on V3 Control Sta-
tions
Expand ADS GUI to Fill the Entire Monitor
See
,
Expanding the ADS GUI to Fill the Monitor (V3 and V4 Control Stations)
.
Nothing to do as Z400 uses
Xorg server
for the graphics management which
auto-detects - see
,
lution Configuration on V4 Control Sta-
tions
.
V3 Control Stations
(U20 - Xsun server)
V4 Control Stations
(Z400 - Xorg server)