Job Card D/R A227 - Rotation Potentiometer
6
PROCEDURE
6-1
Disassembly of the Rotation Potentiometer
1. Remove the Rotation Board supporting plate (2.5 mm allen wrench).
Note:
In the following step, take care not to lose the tension spring (3) when loosening the screw (2).
2. Remove the screw (2) securing the Potentiometer support (2 mm allen wrench).
3. Unplug the Potentiometer cable (W220) from the Rotation Board (connector J11).
3
2