GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Product history and compatibility
Page no. 61
Chapter 3
Product-historyDS.fm
3-1-9
Nephtys M414 (Penduick)
•
New and improved Gantry Boards, as follows:
-
Gantry CPU Board with NSC.1.2.15 (5233827-8)
-
Lift Board with HC12: V5.16.2 (2373567-11)
-
Rotation Board with HC12: V5.12.4 (2375488-13)
-
Compression Board with HC12: V5.12.1 (2373563-8)
•
New Generator CPU Board with 4.02 CMN.
•
New PDU board (2365871-8), with unchanged software, but contains a different F5 fuse (5344386-2)
and a different transformer. This new PDU board is fully backward compatible with IB systems, and
must be used to replace earlier PDU boards.
•
New ADS Software 53.40 (5261381-6).
•
New Dione 5 form factor IDC (5324091).
•
New IDC Software 10.0.5 (5324092) to support the Dione 5 form factor IDC.
•
New Strator Cable W403 (2381701-2) with corrected mA return cable connection on XP2 connector.
This cable is fully backward compatible with IB systems.
•
New Upper Compresion Arm cover (2401145-2), which is fully backward compatible with IB systems.
When changing an Upper Compression Arm cover, always order 2401145-2.
•
Introduction of a new V2 50 N Rotation Clutch, which is only physically compatible with the Rotation
Board 2375488-12 or later.
Note:
Since no more V1 Rotation Clutches exist, the
Spare Part Kit Clutch 2345968-4 and Rotation
Board 2375488-13 (5265895-2)
has been introduced for systems where both the Rotation Board
and Clutch need changing, replacing
Spare Part Kit Clutch 2345968-3 and Rotation Board
2375488-12 (5265895)
.
•
New Mavig Radiation Screen (5352284), which is less high and slightly thinner than the previous
Lamerpax Radiation Screen. The Mavig Radiation Screen is compatible with V2, V3 and V4 Control
Stations.
•
Introduction of a new spare part, Rotation Potentiometer and Support (2372281-2) replacing the
Rotation Potentiometer spare part, which is fully backward compatible. As a result the entire Rotation
Potentiometer and Support are changed.
•
An optional 3 MP 21" Color LCD Monitor (5370904).
•
A new V3 "Penduick" Control Station exists with minor modifactions from the previous V3 Control
Station. See
for more information.
3-1-10 Nephtys M415 (Z400)
•
A new V4 Control Station exists with major modifactions compared to the V3 Control Station. See
for more information.
•
New ADS Software 54.10 (5261381-7) to support the new HP Z400 computer used in the V4 Control
Station.