GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 879
Chapter 9
ERR-SUB-A-003.fm
3-3
Error Codes from 120/yyy: Compression
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card
120/001
WARNING
LIFT UP BUTTON
STILL PRESSED
UIF Still Pressed [i] (for i =1 - 11) = Occurs when one
of the 11 buttons on compression node is still
pressed:
- Lift up (V1, V2)
- Lift down (V1, V2)
- Rotation Right (V1, V2)
- Rotation Left (V1, V2)
- Preset CC
- MLO
- Preset -MLO
[i] indicates the nature of the button. Right or left is
not indicated.
See
to fix the problem.
A1201
120/002
WARNING
LIFT UP BUTTON
STILL PRESSED
120/003
WARNING
LIFT DOWN BUT-
TON STILL
PRESSED
120/004
WARNING
LIFT DOWN BUT-
TON STILL
PRESSED
120/005
WARNING
ROTATION BUT-
TON STILL
PRESSED
120/006
WARNING
ROTATION BUT-
TON STILL
PRESSED
120/007
WARNING
ROTATION BUT-
TON STILL
PRESSED
120/008
WARNING
ROTATION BUT-
TON STILL
PRESSED
120/009
WARNING
CC BUTTON STILL
PRESSED
120/010
WARNING
+MLO BUTTON
STILL PRESSED
120/011
WARNING
-MLO BUTTON
STILL PRESSED
121/001
DEBUG
ONLY
COMPRESSION
SOFTWARE
ERROR
On every command sent by the PTB, the IPMS
checks whether the command identifier is part of the
list of commands to be taken into account by IPMS. If
the command is not recognized, it is ignored, and
“CommandIdError” is returned to the PTB.
1. Change the Compression Board.
2. Perform a calibration of the Compression Force
Sensor.
3. Perform a Compression Thickness calibration.
D/R A250
CAL A048
CAL A050