Chapter 9
Page no. 1750
JC-CHK-A-011.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card CHK A011 - Check kVp settings
8
PROCEDURE - RADCAL 4083 KVP METER
8-1
kVp meter calibration
•
Note the meter identification number and date of most recent calibration:
8-2
Position the meter
1. Ensure that the Radiation Protection Plate is 155 mm (6.10 inches) below the bottom of the Tube
Head.
2. Use the Senographe light centering device to center the kV peak meter target within the X-ray zone.
kVp meter
identification number
Date of calibration:
(dd/mm/yy)