GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A414 - Power Line Filter Box
Page no. 1619
Chapter 9
JC-DR-A-414.fm
particular customer. Disconnect the room and door cables (1) from the Generator Relay board.
7. Disconnect the Line Supply Cables (2) from the Power Line Filter.
8. Carefully disconnect the data cable (3) from the Generator Relay board. Ensure that you do not move
it too much so it is easier to re-connect it later on.
! Notice:
Disconnect with caution to avoid damage to the connectors and cables.
9. Remove the four screws (4) that secure the Generator Relay board (4 mm allen wrench) to the Gen-
erator chassis, and carefully remove the Generator Relay board.
10. Put the Generator Relay board in an anti-static bag in a safe place.
11. Disconnect the earth cable lablelled
FILTRE
(5) from the Power Line Filter (4 mm open ended
wrench).
12. Disconnect the allen screw (6) from the left side of the Power Line Filter (3 mm allen wrench).
13. Disconnect the two allen screws (7) from the right side of the Power Line Filter (4 mm allen wrench).
5