GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A331 - UPS
Page no. 1527
Chapter 9
JC-DR-A-331.fm
7. The battery plug to the UPS is not connected at delivery time. Firmly connect the battery plug (7) to
the battery on the newly installed UPS as shown below.
8
COMPLETION
Power up the Senographe system, and perform a UPS functional check according to
Job Card CHK A003 - UPS Batteries
Before sending the old UPS for recycling, ensure that the plug on the UPS has been disconnected from
the battery as shown below).
7
UPS battery disconnected at delivery time
Connect UPS battery as shown below
Old UPS from V1/V2 Control Stations
Old UPS from V3/V4 Control Stations
Plug
Disconnected