GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A210 - Gantry cover replacement guide
Page no. 1211
Chapter 9
JC-DR-A-210.fm
3-9
Detector rework
Securing holes must be added to the Detector assembly on earlier systems to provide anchor points for
the newer covers.
1. Remove the Detector by following the instructions of
Job Card D/R A141 - Image Receptor/Digital
2. Drill a 4 mm countersink hole at 90° to the exterior face through the mounting lug on each side of the
assembly:
-
Position; 16.7 mm behind the existing hole, as shown below.
-
Diameter: 4 ±0,1mm.
-
Countersink to 7±0,5 mm on the exterior face.
16.7 mm