Chapter 9
Page no. 888
ERR-SUB-A-003.fm
GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
122/041
WARNING
VALID LINE WAS
ACTIVE ON COM-
PRESSION NODE
The valid motion line is active at power-up.
1. Change the Gantry CPU Board (see
2. Perform a Collimator Calibration (see
CAL A029 - Collimator Format Calibration
).
3. Perform a Collimation Check (see
D/R A190
CAL A029
ELE A028
122/042
WARNING
COMPRESSION
WAS STOPPED
WITHOUT STOP
COMMAND
The compression node stopped the motion by itself
because the button/pedal responsible for the motion
was released and no stop order was received before
the timeout.
1. Change the Gantry CPU Board (see
2. Perform a Collimator Calibration (see
CAL A029 - Collimator Format Calibration
).
3. Perform a Collimation Check (see
D/R A190
CAL A029
ELE A028
122/043
HALT
COMPRESSION
FORCE SENSOR
FAILURE
There is a compression force sensor defect. No fre-
quency signal is detected from the compression
force sensor. IPMS_SRS_SAR_013 = In case of
defect on force sensor, the IPMS stops, and forbids
the motion, and sends an error message
IPMS_SRS_ERR_014.
Check cabling between paddle holder and compres-
sion node.
This error can be due to a too high reverse force
applied on the paddle. If the reverse force is too sen-
sitive, the Faulty FRU is the Compression trolley kit.
A1213
122/044
DEBUG
ONLY
COMPRESSION
INVALID
RX_OPTICAL_FOR
K1_ERROR
Ignore this error. If the problem persists, contact your
OLC.
122/045
DEBUG
ONLY
COMPRESSION
INVALID
RX_OPTICAL_FOR
K2_ERROR
Ignore this error. If the problem persists, contact your
OLC.
122/046
DEBUG
ONLY
COMPRESSION
INVALID
LIFT_DOWN_PEDA
L_ERROR
Ignore this error. If the problem persists, contact your
OLC.
Code
Category
Text in Log File or
on LCD
Possible Cause/Recommended Action
Job Card