Job Card D/R A318 - LCD Monitor
7-2-2
Reassembly - 21" 3MP Monitors
If this intervention concerns the replacement of the monitor execute all the following steps, otherwise go
directly to step 3.
1. Remove the four screws (1 and 2) securing the screen plate to the old monitor. The screen plate and
four screws are used for fixing the new monitor to the rotative arm.
2. Place the new monitor on a protected surface. Fix the screen plate to the monitor using washers and
two 20 mm screws (1) on the top, and two 16 mm screws (2) on the bottom.
3. Carefully slide the monitor on to the support at the end of the rotative arm.
2
1
1