Job Card ELE A008 - Configuring the 21-inch 3MP Monitor
Page no. 549
Chapter 7
JC-ELE-A-008.fm
Job Card ELE A008 - Configuring the 21-inch 3MP Monitor
Chapter 7
1
SUPPLIES
None
2
TOOLS
•
(P/N 5175614)
•
A calibrated Luminance Meter that permits measuring the luminance of the monitor in contact (Direct)
mode as illustrated below.
•
If the
(P/N 2319223) is used, it must be used with the foam ring attached,
and must be configured in
Direct
mode.
•
The
Away/Direct
button toggles between away and direct modes. To select the
Direct
mode, keep
pressing the
Away/Direct
button until
D
is displayed on the lower left corner of the Spot Meter
screen.
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 40 minutes
Remove canon and place foam ring
Luminance Meter
ATTENTION:
1. When putting the Luminance Meter into
contact with the monitor screen, do not apply
excessive pressure.
2. When measuring the luminance value of
different parts of the screen, ensure that you
apply the same level of pressure on each dif-
ferent part of the screen.
Foam Ring
Direct
mode selected
Away/Direct
mode
button selector