GE Healthcare
Senographe DS
Revision 1
Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A043 - AWS Monitor Calibration
Page no. 737
Chapter 7
JC-ELE-A-043.fm
7-6
Calibration
7-6-1
18" Monitor and V3/V4 Control Station
1. The SMPTE pattern is already displayed on the AWS Monitor. If not, refer to
to display
the SMPTE pattern.
2. Press the
MENU
button to access the OSD main menu, then keep pressing the
MENU
button until
Auto functions
is selected. Press the
UP
button to enter the
Auto functions
menu. Press the
UP
but-
ton again to perform an auto-calibration.
The auto-calibration takes place.
3. Press the
MENU
button to access the OSD main menu, then keep pressing the
MENU
button until
Service level 2
is selected.
4. Press the
UP
button once, then immediately press and hold the
DOWN
button until the
Service level
2
menu re-appears as shown below. Keep pressing the
MENU
button until
User Settings
is selected.