Job card D/R A252 - Compression Trolley Kit
7
COMPLETION
•
Check compression motion and keypad.
•
Carry out the following calibration procedures: :
Calibrations - Checks
Job Card
Gantry
Compression Force Sensor
Job Card CAL A048 - Calibration of Compression Force Sensor
Compression Thickness
Job Card CAL A050 - Compression Thickness Calibration
If Compression Thickness not OK, repeat calibrations from Compression Force Sensor
AOP/SNR Check
Job Card ELE A038 - AOP Mode and SNR Check
If AOP/SNR Test not OK, repeat all calibrations