86
ENGLISH
SpO
2
panel. However, once the interference disappears, or probe / finger is off, then icon “
” will disappear.
3. The alarm will be triggered if the SpO
2
measurements exceeds the limits totally three times in one minute despite the
Sat-seconds limit is not reached yet.
Below is the related alarm information for Nellcor SpO
2
module.
No.
Event
Alarm level
Remark
1
SpO
2
INOP
High priority alarm
Such alarm will not display on alarm information
area, but it will be recorded as alarm event which
can be reviewed on “Alarm event”. these events
are mostly unrecoverable faults.
2
SpO
2
firmware
error
3
Communication error
4
SpO
2
faulty sensor
5
SpO
2
hardware error
6
SpO
2
module error
7
SpO
2
sensor error
It means the SpO
2
sensor identifications fails.
8
SpO
2
sensor disconnected
Medium priority
alarm
The sensor cable or extension cable is not
connected to the monitor.
9
SpO
2
probe off
It means the sensor cable is connected, but the
probe is out of the measuring site.
Chapter 11 Monitoring Temperature
11.1 Introduction
The body temperature is monitored by direct measuring mode with the temperature sensor of thermistor type. Very small
amount of constant current is applied to the temperature sensor to avoid self-heating, the voltage across the thermistor
is measured, and further converted to the temperature reading according to the temperature-resistance characteristic
for a specific type of thermistor. The temperature measuring circuit does the self-testing periodically to prevent false
reading when there is failure in hardware.
You can simultaneously monitor two temperature sites using the device, for specific model, only one channel temperature
is available.
11.2 Safety Information
Verify that the probe detecting function works correctly before monitoring. Unplug the temperature probe cable
from the T1 or T2 connector, and the monitor can display the message [T1 Sensor Off] or [T2 Sensor Off] and give
alarm tones correctly.
Make correct choice in setup menu for “KRK” and “YSI” series of temperature sensor used, since they have different
temperature-resistance characteristic (KRK thermistor has 10.000K ohms @25
o
C, YSI thermistor has 2.252K ohms
@25
o
C). Mismatching will cause wrong temperature reading or even out of range.
11.3 Making a TEMP Measurement
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...