![background image](http://html1.mh-extra.com/html/gima/35306/35306_manual_2228144067.webp)
67
ENGLISH
Heart Rate Panel:
Heart rate value
Heart rate label and unit
Heart beat symbol
High and low alarm limit
setting for Heart rate
Arrhythmia symbol
“HR”:
heart rate. The 64 on the right is the heart rate measured.
“bpm”:
the unit of heart rate, it means “beats per minute”.
“
“:
the heart-beat symbol, blinks corresponding to the R wave of ECG waveform.
“180/40”:
high and low alarm limit setting for heart rate.
“
“: Arrhythmia symbol. If the ARR function is set as ON, then this symbol will be displayed.
7.5 Changing ECG Settings
Select “Menu”
→
“ECG” to enter into ECG related setting.
Note: according to the different function configuration, the ECG parameter setting window may be different, please refer
to the monitor in your hand.
Speed:
the ECG waveform sweeping speed, 4 options: 6.25mm/s, 12.5mm/s, 25mm/s and 50mm/s. The factory
default is 25 mm/s.
Filter
: ECG filtering mode, 3 options: MONI, DIAG, and OPER.
DIAG
: extended bandwidth to provide the ECG waveforms with the quality of diagnosis level.
MON
I
: normal bandwidth to provide noise free ECG waveforms for effective monitoring purpose.
OPER
: narrow bandwidth to deeply suppress the interference from electro-surgical unit during operation.
The default mode is “MONI”.
Notes:
1. For different ECG related setting window, and when the ECG filtering mode is set as “OPER”, then items like
“Wires”, “Lead” and “Pacer” are in grey and nonadjustable. The “Wires” is fixed to be “3 lead wires”, the “Lead” is
fixed to be lead “II”, and the “Pacer” is fixed to be unselectable.
2. If the amplitude of an ECG waveform is too large, the peak of the waveform might not be displayed. In this case,
you should change the waveform gain properly.
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...