68
ENGLISH
Gain:
The ECG gain, options: x1/8, x1/4, x 1/2, x 1, x 2, x 4 and Auto. The “Auto” is for automatic gain control. The
default is “x1”.
x1/8: 1/8 of the base gain; x1/4: 1/4 of the base gain;
x1/2: 1/2 of the base gain; x1: the base gain;
x2: twice of the base gain; x4: 4 times of the base gain;
Auto: auto gain control; The base gain is 10 mm/mV.
Notch:
only in “DIAG” filtering mode the notch filter can be chosen.
3 options: OFF, 50Hz, 60Hz. The factory default
is 50Hz.
Wires:
The default is “5-lead”, it is necessary to select ECG Lead with “5-lead” setting to obtain all ECG signals
including Lead I, II, III, aVR, aVL, aVF, and V. The user can also choose “3-lead” setting. When “3-lead” is chosen, only
electrodes of R/RA, L/LA, and F/LL are used for detecting ECG signals, and the user can only select ECG signal of Lead
I, or II, or III, the default Lead selection is Lead “II”.
Note: if your monitor is configured with 12-lead ECG function, then you can set “Wires” as “5-lead” or “12-lead”
to obtain ECG signals. When “12-lead” is chosen, then the waveform of lead I, II, III, aVR, aVL, aVF, V1~V6 will be
displayed on the All ECG Trace view.
Cal 1mV:
Generate the 1mV calibration signal internally. This signal is used to self-test the signal amplitude of the
monitor. It must be unselected during normal operation. The default is unselected.
Grid:
The display switch of the grid lines on the background of ECG waveform. The default is OFF.
Pacer:
Enable the cardiac pacemaker pulse detection, the default is unselected. When the “Pacer” is selected, the
function of pacemaker pulse detection will be effective. A mark “ “ will be overlapped on the ECG waveform (as
shown in the following figure.) if the pacemaker pulse is detected while the patient wears a cardiac pacemaker.
Note:
pacemaker pulse inhibition function is always effective for Heart Rate calculation whether you enable or disable the
function of cardiac pacemaker pulse detection or not.
PARM:
to enter into HR alarm setting. See Section
“Understanding the Alarm Setting”.
7.6 About Arrhythmia Detection and Template Learning
ARR:
ECG arrhythmia analysis
ARR switch:
the switch of ECG arrhythmia detection. The default is ON. This switch does not affect the alarm function
for 5 ARR events (ECG Brady, ECG Tachy, ECG Arrest, VE Run and SVE Run). That’s to say, the detection and alarm
for 5 ARR events is fixed to be ON, and nonadjustable.
Extra ARR Learning:
indication of ARR learning status or activating ARR learning manually.
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...