Gain:
the CO
2
waveform gain. 4 options: X1/2, X1, X2 and X4. The default is X1 for adult and pediatric patient, and
X2 for neonate patient.
X1 waveform scale with base gain X1/2 half scale size of the base gain
X2 twice scale size of the base gain X4 four times scale size of the base gain
Speed:
Respiration waveform sweeping speed of, 2 options 6.25mm/s and 12.5 mm/s. The default is 12.5 mm/s.
Apnea(s):
The timeout setting for apnea alarm (in second).
(1). When AG monitoring is on:
It can be set as any number from 20 to 60 seconds, the step is 1 second. Icon “
“ displays on the
upper right corner of Respiration Panel; When the device has not detected any breathing signal for the specified
time, the “Apnea” alarm prompts on, and alarm sound will be activated. If it is set as OFF, the icon “
“
displays on lower left corner of the Respiration Panel. The default is 20s.
(2). When AG monitoring is off:
refer to section
Changing RESP Setting
.
Source:
Respiration signal source. This item is fixed to be “AG” if AG monitoring function is selected. Otherwise, the
source will be obtained from thoracic impedance measurement from ECG module.
Zeroing:
press it to perform zero resetting.
Default:
resume to the factory default value.
CO
2
Unit:
the unit of CO
2
concentration. 3 options: kPa, mmHg and vol%. 1kPa=7.5mmHg. CO
2
(mmHg) = (CO
2
concentration (vol%) ) * (atmospheric pressure (kPa) ) * 7.5.
Barometric:
the detected atmospheric pressure, which can be set at the same time. The setting range is from
1~999(%), the default is 100(%).
O
2
density:
Oxygen concentration setting.
: the switch for displaying waveform of AA1 and AA2. Select it means the AA1 and
AA2 waveform will be displayed on the screen during monitoring.
PARM:
to enter into Anaesthesia Gas alarm setting. See Section
“Understanding the Alarm Setting”.
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...