141
ENGLISH
Chapter 21 Other Functions
21.1 System Information
Select “Menu”
→
“System Info”, then the “System Info” window can be entered, which including release version,
software version, hardware version and product ID.
Product ID:
set by manufacturer for traceability.
For some monitors, it may also show embedded module information:
Embedded Module:
other extended modules except the built-in module (ECG/ SpO
2
/ NIBP/ RESP/TEMP), such as CSM
and CO
2
module.
The current installed module information will also be displayed, for example, “Plug-and-Play modules L: CSM-IoC R:
EtCO2”.
21.2 Nurse Call Settings (Optional)
Output level:
the output level from nurse call connector, “High level” and “Low level” for option.
Duration:
two output modes for optional: “Continuous” and “Pulse”. The continuous mode of output means the
nurse call signal will keep until the alarm disappears. The pulse mode means the output nurse call signal lasts for
one second.
Trigger Condition:
3 kinds of alarm sources can trig the nurse call: high level alarm, medium level alarm and low
level alarm.
Do not rely exclusively on the nurse call system for alarm notification. Remember that the most reliable alarm
notification combines audible and visual alarm indications with the patient’s clinical condition.
Chapter 22 Battery
22.1 Overview
The monitor has built-in chargeable battery, when the monitor is disconnected from the external AC power supply, then
it will be powered by the built-in battery. When the monitor is supplied by external AC power, then the battery will be
charged. Generally, it’s unnecessary to maintain the battery when using under normal working condition.
When the monitor is powered by built-in battery, and battery is in low voltage, then technical alarm will be triggered, and
message “Low battery” pops up. At this moment, the user should connect the external power immediately to make sure
the monitor continue working.
On-screen battery symbols indicate the battery status as follows:
AC Power supply and the battery is fully charged.
The red exclamatory mark blinks means battery voltage will be used out and needs to
charged. Alarm information area will show “Low Battery” message.
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...