review the current record on C.O. waveform area, or the user can press “
“ and “
“ to review the records.
Note: The current C.O. measurement result will be displayed in the table “Examin. records” and saved in history records
as well. When the device is shutdown, the content listed in the table “Examin. records” will be lost, but the history records
still exists in the database. The history record includes patient ID, measuring time, TI, TB, C.O., C.I., and BT waveform as
well.
Average C.I.(l/min/m
2
):
the calculated average C.I. for the past examination records.
Average C.O.(l/min):
the calculated average C.O. for the past examination records.
You can press “
” button to delete the record if the tolerance is too large.
14.6 Factors Affecting C.O.Measurement
Heart
A. Preload: the blood volume in the ventricle at the end of diastolic phase, which is related with the volume of
venous return and the residual blood. Normally the pulmonary capillary wedge pressure (PCWP, its normal value is
6-12mmHg, or 0.8-1.6kPa) is regarded as the reliable indication for the preload in right ventricle.
B. Afterload: the resistance against blood ejection.
C. The myocardial contraction and its harmonization, the heart compliance and heart rate.
Monitor
The type of catheter, the correction constant of this catheter, the temperature, volume and speed of injection.
Temperature of injectate solution,
Volume of injectate solution,
Patient’s baseline blood temperature,
C.O. record list area
C.O. waveform area
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...