55
ENGLISH
Medium
ECG/SpO
2
Lead(s) off, SpO
2
Probe off, SpO
2
value exceeds measuring range
CO
2
CO
2
sensor off, need to perform zero resetting, CO
2
over detect value, check air
adapter, check sampling tube, CO
2
zeroing failed.
IBP
IBP1 Probe off, IBP2 Probe off, IBP3 Probe off, IBP4 Probe off
CSM
Electrodes off
Temp
Temp probe off
AG
Sidestream: No sampling line, Sampling line block
Mainstream: No adapter, Required to change adapter
Low
NIBP
Self-test failed, System error, Over time, Signal weak, Cuff error, Air leakage, Pressure
error, Out of range, Too much motion, Over pressure, Signal saturation, Detected air
leakage, BP exceeds measuring range,
TEMP
Temperature self-detect failed, TEMP1/TEMP2 exceed(s) measuring range
SpO2
No pulsation
System
AC-DC low voltage
Related to 3 levels alarm sources, medical and nursing staff should have different response to deal with potential dangers,
the detailed demands are as follows:
1. High priority alarm: medical and nursing staff should response immediately.
2. Medium priority alarm: medical and nursing staff should response quickly.
3. Low priority alarm: medical and nursing staff should response as soon as possible.
NOTE:
Some models may have only Medium and Low alarm priority because of the different configuration.
6.3 Alarm Indicators
When an alarm occurs, the monitor will indicate it to the user through visual or audible alarm indications.
6.3.1 Alarm lamp
Lamp color
Alarm level
Red flashing
High priority alarm
Yellow flashing
Medium priority alarm
Yellow
Low priority alarm
Green
Normal
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...