78
ENGLISH
“PR 62”:
pulse rate value when taking blood pressure measurement.
“Manu”
: the icon of NIBP measurement mode. There are 3 modes: “Manual”, “Auto” and “STAT”. When in “AUTO”
mode, a count-down timer is displayed as well.
9.7 Changing NIBP Settings
Select “Menu”
→
“NIBP” to enter into NIBP related setting.
Mode:
“MANU”, “AUTO”, “STAT” and “Customized Multi-cycle” can be selected. The default is “MANU”.
In “MANU” mode, press the NIBP measure key “
” manually can start or stop NIBP measurement.
In “AUTO” mode, the device repeats NIBP measurement automatically with the set time interval. In this mode,
the manual interference still works.
In “STAT” mode (only used for adult), press the NIBP measure key “
”, the device will do NIBP
measurement again and again. The device will not stop making measurement until the measuring time is over
5 minutes or the operator stops it manually.
Customized Multi-cycle
: move the cursor on “Customized Multi-cycle” and click “OK” to enable this function. The
user can customize the related parameters: Phase, time cycle (the time interval between two measurements) and
repeats. There are 5 phases: A, B, C, D and E. The user can set the time cycle and repeats for Phase A to Phase E.
Time cycle
: 1min, 2min, 3min, 4min, 5min, 10min, 15min, 20min, 25min, 30min, 35min, 40min, 45min, 50min,
55min, 1h, 1.5h, 2h, 2.5h, 3h, 3.5h and 4h for optional.
Repeats
: OFF, 1, 2, ...9 and 10 for optional.
For example, firstly, the monitor enter into Phase A (making NIBP measurement once every 5 minutes, and repeat once
only); Secondly, entering into Phase B (making NIBP measurement once every 10 minutes, and repeat once only);
Thirdly, entering into Phase C (making NIBP measurement once every 20 minutes, and repeat 2 times); Fourthly,
entering into Phase D (making NIBP measurement once every 30 minutes, and repeat 5 times); Lastly, entering into
Phase E (making NIBP measurement once every 60 minutes, and repeat once only). In the period of 30 minutes, if
the NIBP measurement is less than 6 times and NIBP measurement mode is not changed, then the monitor will start
to make NIBP measurement from Phase A to E automatically.
Caution:
STAT mode can only be used for adult.
Notes:
When changing patient, the default NIBP measuring mode is “Manual”.
For “Adult”, if the NIBP measuring mode is set as “STAT”, then the monitor will not save this setting when you shutdown
the monitor. That’s to say, when re-starting the monitor, the NIBP measuring mode for Adult is “Manual”.
For all patient types, if the NIBP measuring mode is set as “Manual”, or “Auto”, or “Customized Multi-cycle”, then the
monitor will save this setting when you shutdown the monitor.
Cycle:
it can be set only in “AUTO” mode. Cycle means the time interval between measurements when the measuring
mode is set to Auto. The Cycle options are: 1min, 2min, 3min, 4min, 5min, 6min, 7min, 8min, 9min, 10min, 15min,
20min, 25min, 30min, 35min, 40min, 45min, 50min, 55min, 1h, 1.5h, 2h, 2.5h, 3h, 3.5h, 4h, 4.5h, 5h, 5.5h, 6h, 6.5h,
7h, 7.5h and 8h. that’s, the range is from 1 minute to 10 minutes, step is 1 minute; for range from 10minutes~1hour,
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...