![background image](http://html1.mh-extra.com/html/gima/35306/35306_manual_2228144034.webp)
34
ENGLISH
Please contact the local dealer or our company in case of any problems. We will offer the best solution for your satisfaction.
3.1.2 Environmental Requirements
The operating environment of the equipment must meet the requirements specified in this manual. Otherwise,
unexpected consequences, e.g. damage to the equipment, could result.
The environment where the equipment is used shall be reasonably free from noise, vibration, dust, corrosive, flammable
and explosive substances. If the equipment is installed in a cabinet, sufficient space in front and behind shall be left for
convenient operation, maintenance and repair. Moreover, to maintain good ventilation, the equipment shall be at least 2
inches (5cm) away from around the cabinet.
When the equipment is moved from one place to another, condensation may occur as a result of temperature or humidity
difference. In this case, never start the system before the condensation disappears
.
3.2 Getting Started
3.2.1 Connecting to Power Source
1. Using AC Power Source
Make sure that the AC power supply is (100-240) VAC, 50Hz/60Hz.
Use the power cable provided by the manufacturer. Insert one end of it to the AC power input of the monitor
and the other end to the three-pin outlet of the power source with protected-earth.
To eliminate potential differences, the monitor has a separate connection to the equipotential grounding
system. Connect one end of the provided ground wire to equipotential grounding terminal on the rear of the
monitor, and connect the other end to one point of the equipotential grounding system.
Caution:
1. Ensure that the monitor is grounded correctly.
2. If you have any doubt to the grounding layout and its performance, you must use the built-in battery to power
the monitor.
After the mains power supply has been interrupted while power switch remains in the “on” status and is restored
after a period of time that is longer than 30 seconds, the monitor will run by the last settings when restarting the
monitor.
The monitor is applicable to connect to the public mains power network.
2. Using Battery
The following steps should be followed to install the battery:
Step 1: open the battery cover;
Step 2: pull out the battery cable and connect it to the battery pack;
Step 3: push the battery pack into the battery compartment and lock it;
Step 4: close the battery cover.
Caution:
it’s better to recharge the battery after it is used up, and the charging time should be 12~15 hours long.
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...