ENGLISH
Adjusting Volume
1. Alarm Volume
To set the alarm sound volume.
Step 1: Select “Alarm”
→
“Others”
→
“Alarm Volume”. Its setting range is “1 ~ 10”, the step is 1. The default is 5. “10” is the
maximum volume. (Note: if your monitor is optioned “Spot-check Mode” function, then the setting options for
Alarm Volume is “Spot Check”, “1~10”.)
2. Beep Volume
To set the pulse beep sound.
Step 1: Select “Menu”
→
“System Settings”
→
“General”.
Step 2: In “General” menu, select “Beep Volume”. Its setting range is “OFF”, “0 ~ 10”, the step is 1. The default is 5. If you
want to set the beep volume as “OFF”, you need to enter the password.
Attention:
It’s not recommended to set
it as OFF.
3. Key Tone
To enable or disable the keystroke sound and the touch-screen sound.
4. Others
Mode:
monitor working mode selection. See
Operating Modes
.
Freeze type:
to select waveforms to be frozen on screen. The options are “All Waves”, “EEG” and “ECG Waves”. When
“ECG Waves” is selected, the system only freezes ECG waveforms. When “EEG” is selected, the system only freezes
EEG and EEG waveform. When “All Waves” is selected, the system freezes all waveforms on screen. The default is
“ECG Waves”. On CSM view, it freezes EEG waveform. And when “All Waves” is selected, the EEG waveform will be
priority to be measured.
Param font:
to select parameters’ digit font, there are 2 options: Bold Digit and Serif Digit.
Plethysmogram:
the switch to turn on or off the filled waveform style for plethysmogram display. When it is ON, the
monitor displays filled waveform for plethysmogram, otherwise, displays in simple curve style. The default is OFF.
Note: this function is not available for ECG waveform.
Beep volume
: to set the pulse beep sound. Its setting range is “OFF”, “1 ~ 10”, the step is 1. The default is 5. Attention:
It’s not recommended to set it as OFF.
Pitch Tone:
to choose the style of pitch tone, the pitch tone means the tone of pulse beep (from ECG or Oximetry)
changes when the measured SpO
2
value changes, the higher SpO
2
value, the sharper beep sound (high-pitch). Two
options: “Mode I” and “Mode II”. The difference between Mode I and Mode II is the tone frequency at the same SpO
2
value. The factory default setting is “Mode I”.
Screen brightness:
to set the brightness of LCD backlight, 3 levels (1, 2and 3) for optional (note: for some specific
module, only 2 levels for optional, that’s “1” and “2”), the factory default is level 1. Level 1 is darker. (Note: this
function is only available for Series III and Series IV.)
Key backlight:
to turn on/off key backlight for the device, the factory default is “On”. (Note: this function is only
available for Series III and Series IV.)
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...