5. TD 0.5: Temperature difference, namely the difference between temperature 1 and temperature 2.
11.5 Changing TEMP Settings
Select “Menu”
→
“TEMP” to enter into TEMP related setting.
Temp Unit:
Temperature unit. The default is
℃
(Celsius) and can be set as
℉
(Fahrenheit), the default is “
℃
”.
Probe type:
KRK and YSI can be selected. KRK and YSI are two different type. If the used type and the set type is
different, then the measured value is invalid.
TD:
the absolute temperature difference. When the temperature difference is higher than preset value, then the
device will activate alarm. If there is only one temperature , the
TD
shows “
――
”.
PARM:
to enter into RESP alarm setting. See Section
“Understanding the Alarm Setting”.
Chapter 12 Monitoring IBP
12.1 Introduction
Invasive Blood Pressure (IBP) is a direct measurement of the patient’s arterial or venous blood pressure. A catheter is
used and inserted directly into a vein, artery or other pressure access areas, and is connected to a pressure transducer
for measuring systolic, diastolic and mean blood pressures. The device can choose to monitor 2 or 4 (for specific model)
channels IBP depending on the configuration. The device can display the systolic, diastolic and mean blood pressures and
a waveform for each pressure channel.
12.2 Safety Information
The pressure tube which connects the catheter and the pressure transducer should be straightway without any
tangle.
Use the pressure transducer kit specified in this manual. Never reuse disposable pressure transducers.
Make sure that the applied parts never contact other conductive parts.
To reduce the hazard of burns during high-frequency surgical procedures, ensure that the monitor’s cables and
transducers never come into contact with the high-frequency surgical units.
When using accessories, their operating temperature should be taken into consideration. For details, refer to
instructions for use of accessories.
Use the accessories specified by the manufacturer, when a defibrillator is used.
Improper use of a defibrillator may cause injury to the patient. The user should determine whether to perform
defibrillation or not according to the patient’s condition.
Before defibrillation, the user must ensure both defibrillator and monitor have passed the system test and can be
safely used jointly.
Do Not use the damaged pressure tube and transducer.
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...