![background image](http://html1.mh-extra.com/html/gima/35306/35306_manual_2228144091.webp)
91
ENGLISH
12.4 Understanding the IBP Display
IBP Waveform:
IBPL1
: Parameter label, it means that the left inserted plugin module is the IBP1.
[0 300]: IBP coordinates scale.
Note: for some monitors, it may only show 2 channels of waveform.
IBP Panel:
IBP Alarm information
The IBP alarm information will be displayed on the alarm information area.
Technical alarm: “Probe off” and so on.
Physiological
alarm: “IBPL1 Low SYS “ and so on.
12.5 Changing IBP Settings
Select “Menu”
→
“IBP” to enter into IBP related setting.
Label:
the name of blood pressure to be measured. It has the following options:
ART---artery pressure PA---pulmonary pressure
CVP---central vain pressure RAP---right atrium pressure
LAP---left atrium pressure ICP---intracranial pressure
AUXP1---auxiliary pressure 1 AUXP2---auxiliary pressure 2
AUXP
-- auxiliary pressure: AUXP1/2 can be chosen if the actual measuring pressure is not in the list of ART, PA, CVP,
RAP, LAP, and ICP. The operator can choose the Calculating Mode according to the type of blood pressure. If artery
pressure is monitored, please choose “dynamic”, then the reading out will include Systolic, Diastolic and MAP. If vein
pressure is monitored, please choose “static”, then only the MAP will be displayed.
Calc. Mode
: when the Label is AUXP1 or AUXP2, the calculate mode can be “static” or ”dynamic”.
Filter:
there are two options for the pressure waveform filtering ---12.5Hz and 40Hz, the factory is 12.5Hz.
Avg. time:
the time period for averaging to calculate Mean Pressure. Setting range from 1s to 12s, the factory is 8s.
Unit:
pressure unit. 2 options: mmHg and kPa.
IBP mark
and unit
Measuring posit
i
on
Mean arterial
pressure
Diastolic pressure
Systolic pressure
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...