7.3 Preparing to Monitor ECG
7.3.1 Preparing the Patient and Device
1. Skin preparation
The quality of ECG waveform displayed on the monitor is a direct result of the quality of the electrical signal received at
the electrode. Proper skin preparation is necessary for good signal quality at the electrode. A good signal at the electrode
provides the monitor with valid information for processing the ECG data. To ensure enough electrolyte material on the
skin of patients, you need to moisten the measuring sites with 70% isopropyl Ethanol. This will usually be sufficient for
ECG monitoring for a short time (30 to 60 minutes).
2. Connect the cable to the connector marked with the “ECG” icon on the signal input panel.
3. Place the electrode to the patient according to
Section 7.3.2.
4. Attach the ECG lead wires to the electrode
5. Make sure the monitor is turned on and is ready for monitoring.
6. After starting the monitor, if the electrodes become loose or disconnected during monitoring, the system will display
“
LEAD OFF
” on the screen to alarm the operator.
It might not display ECG waveform when using ECG cable with 3 lead wires while the setting of “Cable” is set as “5”
in the ECG parameter setup menu. Only single channel of ECG signal can be obtained while using 3 lead wires and
the “Cable” is set as “3”, this ECG signal can be selected between Lead I, Lead II and Lead III.
In order to obtain other Leads of the ECG signals, such as aVL, aVR, aVF and V, the ECG cable with 5 lead wires should
be used and the “Cable” should be set to “5”. At this situation, 7 Leads of ECG signal (lead I, II, III, aVL, aVR, aVF, V)
can be obtained and displayed simultaneously.
Note:
If any side-effect such as allergic or itchy reaction is found, remove the electrodes from the patients immediately.
The symbol indicates that the cable and accessories are designed as the “CF” type level for protection against
electric shocks and with defibrillation-proof capability.
7.3.2 ECG Electrodes Placement
Electrode Placement
The ECG leads and their corresponding locations are as follows:
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...