step is 5 minutes; for range from 1 hour~8 hour, step is 0.5hour. When in AUTO mode and this item is set, the
operator must manually press NIBP measure key “
“ for the first time, then the device will start time
counting down. The device will start taking the next blood pressure measurement automatically after finishing
counting down.
Unit:
the pressure unit. “mmHg” and “kPa” can be selected. 1 kPa =7.5 mmHg.
Init. Pressure:
Cuff pressure to be inflated initially. Its range is different depending on patient type.
for adult:
80, 100, 120, 140, 150, 160, 180, 200 ,220, 240,260,280 mmHg;
for pediatric :
80, 100, 120, 140, 160, 180, 200, 210 mmHg;
for neonate:
60-140 mmHg.
Note: if your monitor is configured with SunTech NIBP module, then the initial cuff pressure range for adult is
120~200mmHg, the initial cuff pressure range for pediatric is 80- 200 mmHg, the initial cuff pressure range for
neonate is 60 – 80 mmHg.
PARM:
to enter into NIBP (SYS, DIA, MAP,) alarm setting. See Section
“Understanding the Alarm Setting”.
Verification:
to verify the NIBP pneumatic system, is used to verify the pressure accuracy and check the air leakage
of the pneumatic system, which should be conducted by technicians in the specified environment. It includes
“Verification A”, “Verification B” and “Leakage Check”.
Verification A & Verification B:
the 2 kinds of verification of the pressure accuracy. Press the relevant button to start
Verification A or B. When completed, it is necessary to stop the verification manually by pressing “Stop” button.
System will also stop the verification when “Exit” button pressed.
Leakage Check:
it is used by technicians to perform a leakage inspection for NIBP pneumatic system..
Important:
Verification A:
the monitor will auto inflate to the preset pressure value (depending on the patient type) by pump, then it
close the deflating valve. Comparing the pressure value on device with that on standard pressure manometer, and verify
that whether the pressure accuracy exceeds the rated tolerance range.
The auto inflating pressure value for Adult: 250 mmHg (33.3 kPa)
The auto inflating pressure value for Pediatric:200 mmHg (26.7kPa)
The auto inflating pressure value for Neonate: 125 mmHg (16.7kPa)
Verification B:
the monitor will close the valve, the pressure should be inflated manually. Comparing the pressure value
on device with that on standard pressure manometer, and verify that whether the pressure accuracy exceeds the rated
tolerance range.
Note: if the verification window pops up during verification, you can press “Close” button to exit form it only, but it is
forbidden to exit from by pressing “Display View Key”.
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...