6.3.2 Alarm message
When an alarm occurs, an alarm message will appear in the technical or physiological alarm area. And the alarm message
is shown in different color to match the alarm level (high priority alarm in red, medium priority alarm in yellow, and low
priority alarm in white).
6.3.3 Highlighting Numeric
If an alarm triggered by an alarm limit violation occurs, the numeric of the measurement in alarm will be highlighted.
6.3.4 Audible Alarm Tones
The alarm tone is distinct from the heart beat tone, keystroke tone and pulse tone in frequency. The alarm tones identify
the alarm levels as follows:
High priority alarm: beep+beep+beep+pause+beep+beep+beep
Medium priority alarm: triple beep
Low priority alarm: single beep
When multiple alarms of different levels occur simultaneously, the monitor will select the alarm of the highest
level and give visual and audible alarm indications accordingly.
6.3.5 Alarm Status Symbols:
Apart from the aforementioned alarm indicators, the monitor still uses the following symbols telling the alarm status.
①”
” means the alarm sound is disabled or means
the alarm volume is 0 (alarm is mute), at this time the user should pay
more attention to the patient
;
② “
” means the alarm system keeps on (is on “Alarm activating status”), and when alarm event occurs, or currently
active alarms does not meet the alarm conditions any longer, it displays icon “
”;
③ When alarm event occurs, after pressing alarm reset key, then it displays icon “
”. Move the cursor on this icon
can bring up the edit box of “Alarm volume” setting.
Remark:
Alarm reset key: pressing this key can perform alarm reset for the current activated alarm event (that’s to say,
the audible alarming will be off, but the visual alarming still keeps effective), then icon “
” displays on the upper
right corner of the screen. The monitor can response to a new alarm event during the alarm reset status, that’s to
say, both visual and audible alarming will be effective when there is a new alarm condition, and the icon “
”
displays on the upper right corner of the screen at the same time. Alarm reset is not a toggle operation, pressing this
key once or more times only makes alarm reset.
The description for technical alarm status indicator and physiological alarm indicator
1) Technical alarm status indicator is an auxiliary indicator for “Alarm lamp”. It’s used for indicating whether the system in
technical alarm status or not. When the system in technical alarm status, then the indicator in blue, otherwise, it stays off.
2) When high priority physiological alarm event occurs, the physiological alarm indicator “
” is flashing with red
color; when middle priority physiological alarm event occurs, the indicator is flashing with yellow color; when the low
priority physiological alarm event occurs, the indicator is in constant yellow light; if there is no physiological alarm event,
the indicator is in constant blue light.
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...