106
ENGLISH
15.4 Setting Up the AG Measurement
The following MultiGas means Anaesthetic Gas including.
If your monitor has MultiGas monitoring function, please follow the instructions below to perform MultiGas monitoring
on patient.
15.4.1 Instruction for PHASEIN IRMA Mainstream Multigas Analyzer
The IRMA mainstream multigas probe is intended to be connection to other medical devices for display of real time and
derived monitoring data of CO
2
, N
2
O, O
2
and the anesthetic agents Halothane, Enflurane, Isoflurane, Sevoflurane and
Desflurane.
It is intended to be connected to a patient breathing circuit for monitoring of inspired/expired gases during anaesthesia,
recovery and respiratory care. It may be used in the operating suite, intensive care unit, patient room and emergency
medicine settings for adult, pediatric and infant patients.
It is NOT intended to be used as the only means of monitoring a patient. It shall always be used in combination with other
Patient Monitor and/or professional human judgments of patient condition. The IRMA probe is intended to be used by
trained and authorized health car processionals only.
1. Assembly Instruction for IRMA Mainstream Analyzer
Installation
1. Plug the IRMA connector into the IRMA input of host device and switch the power on.
2. Snap the IRMA probe on the top of the IRMA airway adapter. It will click into place when properly seated. Please
refer to the following figure.
3. A green LED indicates that the IRMA probe is ready for use.
4. Connect IRMA/airway adapter 15mm male connector to the breathing circuit Y-piece.
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...