: the apnea alarm status in RESP alarm setting, refer to the following Section for details.
8.4 Changing RESP Settings
Select “Menu”
→
“RESP” to enter into RESP related setting.
Gain:
Respiration amplification/gain times, 4 options: X1/2, X1, X2, X4. The default is X1 for adult and pediatric
patient, and X2 for neonate patient.
X1 waveform scale with base gain X1/2 half scale size of the base gain
X2 twice scale size of the base gain X4 four times scale size of the base gain
Speed:
Respiration waveform sweeping speed of, 2 options: 6.25mm/s and 12.5 mm/s. The default is 12.5 mm/s.
Apnea:
The timeout setting for apnea alarm (in second). It can be set as any number from 5 to 120 seconds, the
step is 1 second. Icon “
“ displays on the lower left corner of Respiration Panel; When the device has
not detected any breathing signal for the specified time, the “Apnea” alarm prompts on, and alarm sound will be
activated. If it is set as OFF, the icon “
“ displays on lower left corner of the Respiration Panel.
Note: when “Wires” is set as “12-lead” in ECG related settings, then “Apnea” can be set as off, 10, 15, 20, 25, 30, 35
and 40.
Source:
Respiration signal source. This item is fixed to be “CO
2
” if CO
2
monitoring function is selected. Otherwise, the
source will be obtained by measuring the thoracic impedance via ECG electrodes, the signal source can come from
the ECG electrodes defined by “ECG Lead I” and “ECG Lead II”. (Note: WHEN “12-lead” is selected, then the “Source”
is fixed to be “ECG Lead II” and nonadjustable.)
Zero:
press it to perform CO
2
zeroing.
PARM:
to enter into RR alarm setting. See Section
“Understanding the Alarm Setting”.
Default:
resume to the factory default value.
Note: in RESP settings window, CO2 and/or AG Settings can be entered if your monitor is configured with CO2
monitoring and/or AG Monitoring.
:
click it to
turn on or off CO
2
Monitoring. When the CO
2
monitoring is selected, then all
items setting as “RESP” will shift to “CO
2
”.
See Chapter Monitoring Carbon Dioxide (CO
2
)
.
:
click it to
turn on or off AG Monitoring. When the AG monitoring is selected, then all items
setting as “RESP” will shift to “AG”.
See Chapter Monitoring
Anesthetic Gas (AG).
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...