Chapter 8 Monitoring Respiration (RESP)
8.1 Introduction
Respiration is monitored by measuring the impedance across the thorax via electrodes places on chest. When the patient
is breathing or ventilated, the volume of air changes in the lungs, resulting in impedance changes between the electrodes.
The device applies safe high-frequency current through ECG electrodes into body and measures the change of voltage
between the electrodes to reflect the thoracic impedance while ECG monitoring is not affected. Respiration rate (RR) is
calculated from these impedance changes, and the respiration waveform is displayed on the monitor screen.
8.2 Safety Information
When monitoring the patient’s respiration, it is recommended to use the so-called “non-OR” ECG cable which has
no built-in resistors to prevent the energy loss of defibrillator discharge. Otherwise the performance of respiration
monitoring is degraded.
The respiration measurement does not recognize the cause of apneas. It only indicates an alarm if no breath
is detected when a preset time has elapsed since the last detected breath. Therefore, it cannot be used for
diagnostic purpose.
If operating under conditions according to the EMC Standard EN 60601-1-2 (Radiated Immunity 3V/m), field
strengths above 1V/m may cause erroneous measurements at various frequencies. Therefore it is recommended
to avoid the use of electrically radiating equipment in close proximity to the respiration measurement unit.
8.3 Understanding the RESP Display
RESP waveform:
Respiration Panel:
“RR”
: the label of Respiration. “
rpm
” is the unit of the Respiration Rate (respiration per minute). The big font “14”
is the value of Respiration rate.
“
“: Breath symbol. The blinking frequency is the same as the Respiration Rate.
“
“: the high and low alarm limit setting for Respiration Rate.
Label and unit of respiration
Respiratory Sync mark
The high and low alarm limit
setting of RR
Apnea alarm mark
Respiration rate
Summary of Contents for 35306
Page 2: ...2 ENGLISH Patient Monitor Operator s Manual ...
Page 21: ...21 ENGLISH Series III Front view Rear view Left view Right view 11 ...
Page 22: ...22 ENGLISH Series IV Front view Rear view Left view Right view 11 Extended Module ...
Page 171: ...2 ESPAÑOL Monitor de paciente Manual de operador ...
Page 343: ...2 FRANÇAIS Moniteur patient Manuel d utilisation ...
Page 362: ...21 FRANÇAIS Série III Vue de face Vue arrière Vue de gauche Vue de droite 11 ...
Page 519: ...2 ITALIANO Monitor Paziente Manuale d Uso ...
Page 538: ...21 ITALIANO Serie III Vista anteriore Vista posteriore Lato sinistro Lato destro 11 ...